Product
Description and
Indications for Use
Peracidin
Dialyzer Reprocessing Concentrate is a stabilized mixture of hydrogen
peroxide and peroxyacetic acid. Peracidin is indicated for the in
vitro cleaning and disinfecting of hollow fiber dialyzers.
Peracidin is intended for use with automated and manual dialyzer reprocessing
systems that have been validated by the system manufacturer for use
with peracetic acid. Since Peracidin is formulated to be used only
for dialyzer reprocessing, the user should follow the dilution instructions
recommended by the reprocessing system manufacturer for the use of
peracetic acid concentrates intended for dialyzer reprocessing.
Germicide Level of Activity: Peracidin
is a high-level disinfectant when used in a concentration of 3.25%
for a minimum of 11 hours at a temperature at or below 75o
F (24o C).
Reuse Period: After dilution, Peracidin
can be used for a period not to exceed to seven (7) days, provided
the required conditions of the active agent concentrations and temperature
exist based on the monitoring, described on page 4, "Use and Handling"
of Peracidin. DO NOT RELY SOLELY ON DAYS IN USE.
Material Compatibility: When used in
accordance with the labeling for these products, peracetic acid-based
cleaning and disinfecting agents have been shown to be compatible
with the materials commonly used in dialyzers. See the "References"
page for literature related to peracetic acid and material compatibility.
Consult the "Instructions for Use" from the specific dialyzer's
manufacturer if additional information is required concerning material
compatibility.
Precleaning Agent Compatibility: When
Peracidin is used in accordance with "Instructions for use", it
is not necessary to use a precleaning agent. Peracidin can be used
for both cleaning and disinfection of the dialyzer.
| 
|
It is important to
read and understand all of these Instructions For Use
prior to using Peracidin Dialyzer Reprocessing Concentrate. |
Continue to Next
Section
|