Instructions Front Page

Testing Prior to Dialyzer Rinse

Description & Indications for Use

Rinse Procedure

Warnings & Precautions

Residual Testing Following Rinse

Use & Handling of Peracidin

Product Specifications

Dilution & Handling Instructions

Emergency & Additional Info

Dialyzer Handling After Reprocessing

References & Order Info


Product Description and
Indications for Use
Peracidin
Dialyzer Reprocessing Concentrate is a stabilized mixture of hydrogen peroxide and peroxyacetic acid. Peracidin is indicated for the in vitro cleaning and disinfecting of hollow fiber dialyzers. Peracidin is intended for use with automated and manual dialyzer reprocessing systems that have been validated by the system manufacturer for use with peracetic acid. Since Peracidin is formulated to be used only for dialyzer reprocessing, the user should follow the dilution instructions recommended by the reprocessing system manufacturer for the use of peracetic acid concentrates intended for dialyzer reprocessing.

Germicide Level of Activity: Peracidin is a high-level disinfectant when used in a concentration of 3.25% for a minimum of 11 hours at a temperature at or below 75o F (24o C).

Reuse Period: After dilution, Peracidin can be used for a period not to exceed to seven (7) days, provided the required conditions of the active agent concentrations and temperature exist based on the monitoring, described on page 4, "Use and Handling" of Peracidin. DO NOT RELY SOLELY ON DAYS IN USE.

Material Compatibility: When used in accordance with the labeling for these products, peracetic acid-based cleaning and disinfecting agents have been shown to be compatible with the materials commonly used in dialyzers. See the "References" page for literature related to peracetic acid and material compatibility. Consult the "Instructions for Use" from the specific dialyzer's manufacturer if additional information is required concerning material compatibility.

Precleaning Agent Compatibility: When Peracidin is used in accordance with "Instructions for use", it is not necessary to use a precleaning agent. Peracidin can be used for both cleaning and disinfection of the dialyzer.

It is important to read and understand all of these Instructions For Use prior to using Peracidin Dialyzer Reprocessing Concentrate.