Dilution and Handling Instructions

General Considerations:

  • Peracidin is a single-container germicide. After dilution, mark the proper Peracidin solution expiration date on the container holding the diluted Peracidin solution. The expiration date is either 7 days after dilution or the Peracidin concentrate's expiration date (marked on the shipping carton and on the bottle label), whichever is sooner.
  • Replace the Peracidin container into the original shipping carton, to prevent exposure to light. Do not store Peracidin concentrate or solution in direct sun light.
  • Test the concentration yielded by the reprocessing system with the peracetic acid reagent strips (PeraChek Peracetic Acid Potency Test Kit, Product Number 1701-01) following dilution of each new container of Peracidin.


    Specific Instructions for Manual Systems:

  • For cleaning of dialyzers: a 2% solution (1 part Peracidin and 49 parts AAMI-quality water) is recommended. A stronger concentration of Peracidin may adversely affect the cleaning of the dialyzer, rather than improve it.
  • For disinfecting of dialyzers: a 3.5% solution (1 part Peracidin and 28.5 parts AAMI-quality water) is recommended.


    Specific Instructions for Automated Systems

    Dilute Peracidin in accordance with the instructions provided by the system manufacturer for peracetic acid-based dialyzer reprocessing germicides.

  • For the Renatron1 reprocessing system, mix a 21% solution, shaking or stirring gently to avoid layering. This solution may be attained in either of the following ways:
    1. Two (2) liters of Peracidin filled to a total volume of 2 1/2 gallons (9463 ml.) with AAMI-quality water; OR
    2. 800 milliliters of Peracidin filled to a total volume of 1 gallon (3785 ml.) with AAMI-quality water.
  • For the Echo2 reprocessing system, a 2% solution is recommended for dialyzer cleaning and a 3.43% solution is recommended for disinfection. These solutions may be attained in the following ways:
    1. A 2% solution is 1 part Peracidin and 49 parts AAMI-quality water.
      • For one gallon, measure 76 ml. of Peracidin filled to a total volume of 1 gallon (3785 ml.) with AAMI-quality water.
      • For 2 1/2 gallons, measure 190 ml. of Peracidin filled to a total volume of 2 1/2 gallons (9463 ml.) with AAMI-quality water.
    2. A 3.43% solution is 1 part Peracidin and 29 parts AAMI-quality water.
      • For five gallons, measure 650 ml. of Peracidin filled to a total volume of 5 gallon (18,925 ml.) with AAMI-quality water.
      • For 2 1/2 gallons, measure 325 ml. of Peracidin filled to a total volume of 2 1/2 gallons (9463 ml.) with AAMI-quality water.
  • For the Seratronics DRS-43, older models of the DRS-4 use a 33% solution of peracetic acid, while models produced since early 1993 use a 21% solution. Consult and follow the instructions provided in the DRS-4 user's manual.

1 Renatron is a registered trademark of Minntech Corporation.
2 Echo is a registered trademark of Mesa Laboratories, Inc.
3 DRS-4 is a registered trademark of Seratronics, Inc.

It is important to read and understand all of these Instructions For Use prior to using Peracidin
Dialyzer Reprocessing Concentrate.